Dosimetric impact of multileaf collimator leaf width according to sophisticated grade of technique in the IMRT and VMAT planning for pituitary adenoma lesion

Soo-Min Chae, Ki Woong Lee and Seok Hyun Son _

PDF  |  HTML  |  How to cite

Oncotarget. 2016; 7:78119-78126. https://doi.org/10.18632/oncotarget.12974

Metrics: PDF 1468 views  |   HTML 1821 views  |   ?  


Abstract

Soo-Min Chae1, Ki Woong Lee1, Seok Hyun Son2

1Department of Radiation Oncology, Cheju Halla General Hospital, Jeju, Korea

2Department of Radiation Oncology, Incheon St. Mary’s Hospital, College of Medicine, The Catholic University of Korea, Seoul, Korea

Correspondence to:

Seok Hyun Son, email: [email protected]

Keywords: multi-leaf collimator, radiosurgery, intensity-modulated radiotherapy, volumetric modulated arc therapy

Received: August 23, 2016    Accepted: October 14, 2016    Published: October 28, 2016

ABSTRACT

We analyzed the difference in the dosimetric effect between 5-mm and 2.5-mm multileaf collimator (MLC) leaf width according to the sophisticated grades of intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT). Nineteen patients with pituitary adenomas were selected for this study. The treatment plans were performed according to the size of the MLC (5-mm and 2.5-mm MLC), the type of technique (IMRT and VMAT), and the sophisticated grades of each technique (5-field, 9-field, 13-field, 17-field technique in IMRT and 1-arc and 2-arc techniques in VMAT). The downsizing effects of MLC leaf width were analyzed using target volume coverage (TVC), conformity index (CI), dose gradient index (GI), and normal tissue difference 70% isodose line and 50% isodose line. Upon replacing the 5-mm MLC with the 2.5-mm MLC, TVC and CI improved by 1.30% and 1.36%, respectively, in total plans. The TVC and CI improved by 1.68% and 1.67% in IMRT, respectively, and by 0.54% and 0.72% in VMAT, respectively. TVC improved by 2.53%, 1.82%, 1.34%, and 0.94%, and CI also improved by 2.70%, 1.81%, 1.24%, and 0.94%, in 5-field, 9-field, 13-field, and 17-field IMRT, respectively. TVC improved by 0.66% and 0.43%, and CI also improved by 0.93%, and 0.52% in 1-arc and 2-arc VMAT, respectively. Regarding the target coverage, there were dosimetric benefits of a smaller MLC leaf width. However, the downsizing effect of the MLC leaf width decreased with the use of a more precise RT technique and a more sophisticated grade of the same technique.