Diffusion-weighted MRI in the evaluation of the thyroid nodule: Comparison between integrated-shimming EPI and conventional 3D-shimming EPI techniques

Luguang Chen, Peipei Sun, Qiang Hao, Wei Yin, Bing Xu, Chao Ma, Alto Stemmer, Caixia Fu, Minjie Wang and Jianping Lu _

PDF  |  HTML  |  How to cite

Oncotarget. 2018; 9:26209-26216. https://doi.org/10.18632/oncotarget.25279

Metrics: PDF 1197 views  |   HTML 1590 views  |   ?  


Abstract

Luguang Chen1,*, Peipei Sun1,*, Qiang Hao1, Wei Yin1, Bing Xu1, Chao Ma1, Alto Stemmer2, Caixia Fu3, Minjie Wang1 and Jianping Lu1

1Department of Radiology, Changhai Hospital of Shanghai, Shanghai, China

2MR Application Development, Siemens Healthcare GmbH, Erlangen, Germany

3MR Application Development, Siemens Shenzhen Magnetic Resonance Ltd, Shenzhen, China

*These authors contributed equally to this work

Correspondence to:

Jianping Lu, email: [email protected]

Keywords: diffusion-weighted imaging; thyroid nodule; single-shot echo planar imaging; integrated-shimming echo planar imaging; dynamic shimming

Received: November 17, 2017     Accepted: April 07, 2018     Published: May 25, 2018

ABSTRACT

This study aimed to evaluate whether a prototype echo planar imaging sequence with integrated-shimming (iShim-EPI) can improve image quality in the thyroid gland in comparison to 3D-volume shimming echo planar imaging (3D-Shim-EPI), and to compare ADC values derived from iShim-EPI with those of 3D-Shim-EPI. Twenty-one patients with thyroid disease were enrolled and underwent axial DWIs with iShim-EPI and 3D-Shim-EPI using a 3 Tesla magnetic resonance scanner in this prospective study. Both sets of DWI images were evaluated by two independent observers who identified susceptibility and ghost artifacts and evaluated the images’ capacity to detect thyroid nodules using quantitative scores. The ADC values of the thyroid nodules and the normal thyroid gland were measured two times within a 4-week period. The reproducibility was evaluated using the intraclass correlation coefficient (ICC) and Bland-Altman plots. There were significant differences in the image quality scores for susceptibility (2.81 ± 0.37 vs. 1.93 ± 0.29, p < 0.001), ghost artifacts (2.95 ± 0.15 vs. 1.93 ± 0.29, p < 0.001) and the detectability of thyroid nodules (3.00 ± 0.00 vs. 2.55 ± 0.75, p = 0.008) between the iShim-EPI and 3D-Shim-EPI techniques, except for the ADC values of the thyroid nodules (1.607 ± 0.466×10-3 mm2/s vs. 1.561 ± 0.483 × 10-3 mm2/s, p = 0.184) and contralateral normal thyroid gland (1.295 ± 0.340 × 10-3 mm2/s vs.1.279 ± 0.411 × 10-3 mm2/s, p = 0.777). Both techniques demonstrated excellent agreement between the ADC values using the ICC (range, 0.963 to 0.999) and Bland-Altman plots. The iShim-EPI technique demonstrated significantly higher image quality compared with the conventional 3D-Shim-EPI technique, with no significant differences in the ADC values.